Chapter 6 Daily Routine
The external light turns on when the door is open.
The button also controls the inner chamber light when the door is closed.
To turn the internal illumination on and off
u
Press and hold the silver sample chamber button.
Running Experimental Samples
At this point in the daily routine, you can run experimental samples.
n
For information on applying compensation, see
Chapter 9, Applying Fluorescence Compensation
and
Appendix B, Example 9-Color Immunophenotyping Experiment
n
For information on acquiring experimental samples, see
Pausing the System
To conserve bulk fluidics and prolong laser lifetime, you can pause the system.
Important:
While the system is paused, the following functions are disabled:
n
All functions in the Instrument Control, PMT Control, and Trigger panels
n
Initiating the ZE5-EYE process
n
Running the QC process
n
Shutting down Everest Software and the ZE5 Cell Analyzer
n
In the Instrument tools:
o
Cleaning the sample line and probe
o
Unclogging the sample line, probe, and flow cell
o
Swapping QC beads
The software enables these functions when the system is restarted.
To pause the system
u
In the System section on the Acquisition workspace toolbar, click Pause System.
160 | ZE5 Cell Analyzer and Everest Software
Содержание ZE5
Страница 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Страница 2: ......
Страница 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Страница 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Страница 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Страница 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Страница 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Страница 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Страница 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Страница 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Страница 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Страница 125: ...Working with Optical Filter Configurations User Guide 123 ...
Страница 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Страница 173: ...Setting Up the Run List The plate setup displays a layout of the selected media User Guide 171 ...
Страница 177: ...Setting Up the Run List User Guide 175 ...
Страница 191: ...Setting Up the Run List User Guide 189 ...
Страница 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Страница 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Страница 333: ...Decontaminating the System User Guide 331 ...
Страница 380: ...Appendix B Example 9 Color Immunophenotyping Experiment 378 ZE5 Cell Analyzer and Everest Software ...
Страница 384: ...Appendix C Optional External DI Water and Waste Ports 382 ZE5 Cell Analyzer and Everest Software ...
Страница 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Страница 397: ......