Acquisition/Event Issues
Error
Possible causes
Troubleshooting steps
Dramatic change in PMTs for all
channels
Incorrect beads used
Ensure single peak ZE-
Series QC Beads are being
used. ProLine™ Rainbow
Beads or ProLine™
Universal Calibration Beads
do not fluoresce in certain
channels, especially for the
UV and violet lasers.
Flow cell clogged
Run the unclog process. See
.
Flow cell dirty
Follow these steps, testing
after each to determine
whether the problem has
been resolved:
1. Run the cleaning process.
See
2. Run the Decontamination
wizard. See
.
User Guide | 347
Содержание ZE5
Страница 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Страница 2: ......
Страница 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Страница 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Страница 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Страница 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Страница 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Страница 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Страница 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Страница 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Страница 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Страница 125: ...Working with Optical Filter Configurations User Guide 123 ...
Страница 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Страница 173: ...Setting Up the Run List The plate setup displays a layout of the selected media User Guide 171 ...
Страница 177: ...Setting Up the Run List User Guide 175 ...
Страница 191: ...Setting Up the Run List User Guide 189 ...
Страница 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Страница 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Страница 333: ...Decontaminating the System User Guide 331 ...
Страница 380: ...Appendix B Example 9 Color Immunophenotyping Experiment 378 ZE5 Cell Analyzer and Everest Software ...
Страница 384: ...Appendix C Optional External DI Water and Waste Ports 382 ZE5 Cell Analyzer and Everest Software ...
Страница 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Страница 397: ......