Running Quality Control
n
Tools section on the Acquisition workspace toolbar
The QC process utilizes single-peak beads conveniently located on board. The beads contain a
mixture of dyes that can be excited by all lasers installed on the ZE5 Cell Analyzer and that can emit in
every fluorescence channel. The level of QC beads is monitored and displayed along with the fluidics
status information.
The process first builds a QC workspace and begins acquiring beads. It then ensures that the system
can acquire the beads at both 150 events per second and 500 events per second. Then, at 500 events
per second, the system automatically adjusts laser delays and PMT voltages so that the mean
intensity of the bead population falls into data channel 128 K in all fluorescence channels. CV
increases and PMT voltage changes are compared against the QC criteria. Finally, a 5,000-event file
is collected at 150 events per second and voltages and CVs are stored for reporting purposes. These
values can be displayed, based on the date the process was run, in daily QC reports and QC trending
reports.
Important:
While QC is in progress, the following functions are disabled:
n
Returning to the Home window
n
Initiating the ZE5-EYE process
n
Sending the run list to the local instrument (performed in the Analysis workspace)
n
In the Instrument tools:
o
Returning sample probe to home position
o
Cleaning the sample line and probe
o
Unclogging the sample line, probe, and flow cell
o
Swapping QC beads
o
Pausing the sheath fluid and disabling the lasers
n
Opening the sample loader door
The software enables these functions when the QC process completes.
While QC is in progress, click the down arrow to display the QC Details window, which provides real-
time updates on each portion of the process.
User Guide | 157
Содержание ZE5
Страница 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Страница 2: ......
Страница 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Страница 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Страница 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Страница 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Страница 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Страница 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Страница 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Страница 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Страница 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Страница 125: ...Working with Optical Filter Configurations User Guide 123 ...
Страница 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Страница 173: ...Setting Up the Run List The plate setup displays a layout of the selected media User Guide 171 ...
Страница 177: ...Setting Up the Run List User Guide 175 ...
Страница 191: ...Setting Up the Run List User Guide 189 ...
Страница 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Страница 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Страница 333: ...Decontaminating the System User Guide 331 ...
Страница 380: ...Appendix B Example 9 Color Immunophenotyping Experiment 378 ZE5 Cell Analyzer and Everest Software ...
Страница 384: ...Appendix C Optional External DI Water and Waste Ports 382 ZE5 Cell Analyzer and Everest Software ...
Страница 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Страница 397: ......