Acquiring Initial Data
6.
In the FSC/SSC plot, position the gate over the relevant part of the sample population.
7.
In the PMT Control panel, adjust PMT voltages to position the population.
8.
After setting up gates and voltages, click Record to save data for the sample.
This clears existing data, disables PMT adjustment, and starts saving a data file. Data are
acquired until the set limit is reached or until you click Stop; sampling does not proceed
automatically to the next position.
9.
Repeat
through
for other samples as needed.
User Guide | 369
Содержание ZE5
Страница 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Страница 2: ......
Страница 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Страница 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Страница 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Страница 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Страница 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Страница 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Страница 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Страница 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Страница 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Страница 125: ...Working with Optical Filter Configurations User Guide 123 ...
Страница 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Страница 173: ...Setting Up the Run List The plate setup displays a layout of the selected media User Guide 171 ...
Страница 177: ...Setting Up the Run List User Guide 175 ...
Страница 191: ...Setting Up the Run List User Guide 189 ...
Страница 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Страница 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Страница 333: ...Decontaminating the System User Guide 331 ...
Страница 380: ...Appendix B Example 9 Color Immunophenotyping Experiment 378 ZE5 Cell Analyzer and Everest Software ...
Страница 384: ...Appendix C Optional External DI Water and Waste Ports 382 ZE5 Cell Analyzer and Everest Software ...
Страница 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Страница 397: ......