Chapter 6 Daily Routine
2.
Pull the waste bottles out of the instrument.
3.
Unscrew the cap assemblies and remove them from the bottles.
4.
Empty the bottles following lab procedures for liquid biohazardous and chemical waste removal.
5.
Carefully place the cap assemblies onto the emptied waste bottles and tighten the caps.
6.
Place the waste bottles back into the instrument.
7.
Attach the cap assembly tubing to the connectors on the instrument.
Tip:
An audible click indicates that the tubing is connected.
Replacing Sheath Bottles
Each sheath bottle has a single connection to the instrument. Waste and sheath bottle connections
are specifically keyed to their respective ports and cannot be connected to the wrong ports.
Important:
When refilling the sheath fluid bottles:
n
Ensure that the filter and tubing are not contaminated when you remove the cap assemblies from
the sheath bottles.
n
Ensure that fluid reaches no higher than the fill line, as shown in the next figure, so that fluid does
not leak from the air vent port in the cap.
146 | ZE5 Cell Analyzer and Everest Software
Содержание ZE5
Страница 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Страница 2: ......
Страница 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Страница 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Страница 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Страница 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Страница 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Страница 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Страница 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Страница 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Страница 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Страница 125: ...Working with Optical Filter Configurations User Guide 123 ...
Страница 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Страница 173: ...Setting Up the Run List The plate setup displays a layout of the selected media User Guide 171 ...
Страница 177: ...Setting Up the Run List User Guide 175 ...
Страница 191: ...Setting Up the Run List User Guide 189 ...
Страница 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Страница 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Страница 333: ...Decontaminating the System User Guide 331 ...
Страница 380: ...Appendix B Example 9 Color Immunophenotyping Experiment 378 ZE5 Cell Analyzer and Everest Software ...
Страница 384: ...Appendix C Optional External DI Water and Waste Ports 382 ZE5 Cell Analyzer and Everest Software ...
Страница 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Страница 397: ......