Experiment Builder
Item
Function
Opens the media selector so that you can change the type of input
device (for example, switching a 40-tube rack to a 96-well plate).
Note:
Switching to a different media type resets the run list
order. For more information, see
.
Adds a stat tube to the plate in the selected Sample position.
Note:
Stat tubes can be assigned only the Sample position on
the plate.
Removes the selected well(s) or stat tube from the run list.
When set to ON, allows you to quickly change the location of a
sample from a filled well to an empty well.
Note:
The default setting is OFF.
Allows you to assign a name to a well. The FCS file that is saved
for the well includes this name. Displays the assigned name of the
selected Plate Map position.
Table 18. Selected Media items and their functions
Position Types
Below the plate map, the Type setting allows you to designate a sample position or group of sample
positions as Sample, Setup, Wash, or Reagent.
Button
Function
Appears in
plate map as
Designates the position as wash, containing water
or cleaning solution. No data acquisition occurs for
this position type.
Designates the position as a sample. Data
acquisition occurs for this position type.
Designates the position as reagent that can be
added to other wells/tubes. No data acquisition
occurs for this position type.
Table 19. Position Type items and their functions
User Guide | 75
Содержание ZE5
Страница 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Страница 2: ......
Страница 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Страница 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Страница 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Страница 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Страница 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Страница 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Страница 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Страница 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Страница 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Страница 125: ...Working with Optical Filter Configurations User Guide 123 ...
Страница 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Страница 173: ...Setting Up the Run List The plate setup displays a layout of the selected media User Guide 171 ...
Страница 177: ...Setting Up the Run List User Guide 175 ...
Страница 191: ...Setting Up the Run List User Guide 189 ...
Страница 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Страница 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Страница 333: ...Decontaminating the System User Guide 331 ...
Страница 380: ...Appendix B Example 9 Color Immunophenotyping Experiment 378 ZE5 Cell Analyzer and Everest Software ...
Страница 384: ...Appendix C Optional External DI Water and Waste Ports 382 ZE5 Cell Analyzer and Everest Software ...
Страница 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Страница 397: ......