MKF+ MKFT (E3.2) 07/2017
page 90/115
20.5 MKFT (E3.2) technical data
Chamber size
115
240
720
Exterior dimensions
Width, gross (including 18 mm /
0.7 in
for 1 access
port (MKFT 115, 240), 36 mm /
1.4 in
for 2 access
ports (MKFT 720), with plug)
mm /
inch
1000 /
39.37
1135 /
44.69
1615 /
63.58
Height, gross (incl. castors)
mm /
inch
1725 /
67.91
1940 /
76.38
2005 /
78.94
Depth, gross (incl. cable and door handle)
mm /
inch
915 /
36.02
1000/
39.37
1230 /
48.43
Depth, gross (incl. cable and door handle) with op-
tional compressed air dryer
mm /
inch
1085 /
42.72
1170 /
46.06
1400 /
55.12
Depth, gross (incl. cable and door handle) with volt-
age and frequency changer
mm /
inch
1530 /
60.24
1615 /
36.58
1845 /
72.64
Wall clearance rear (minimum)
mm /
inch
300 /
11.81
300 /
11.81
300 /
11.81
Wall clearance rear with optional compressed air
dryer or to set up the voltage and frequency chang-
er (minimum)
mm /
inch
1000 / 39.37 1000 / 39.37 1000 / 39.37
Wall clearance sides (minimum)
mm /
inch
200 /
7.87
200 /
7.87
200 /
7.87
Window width
mm /
inch
288 /
11.34
508 /
19.99
508 /
19.99
Window height
mm /
inch
222 /
8.74
300 /
11.81
300 /
11.81
Doors
Number of doors
1
1
1
Interior dimensions
Width
mm /
inch
600 /
23.62
735 /
28.94
1200 /
47.24
Height
mm /
inch
480 /
18.90
700 /
27.56
1020 /
40.16
Depth
mm /
inch
400 /
15.75
443 /
17.44
600 /
23.62
Interior volume
l /
cu.ft.
115 /
4.06
228 /
8.05
734 /
25.92
Racks
Quantity of racks (regular)
1
1
1
Quantity of racks (max.)
pc.
4
6
11
Maximum load per rack
kg /
lbs.
30 /
66
30 /
66
40 /
88
Maximum permitted total load
kg /
lbs.
60 /
132
70 /
155
160 /
353
Weight
Weight (empty)
kg /
lbs.
330 /
728
415 /
915
635 /
1400
Weight (empty)
with optional compressed air dryer
kg /
lbs.
345 /
761
430 /
948
650 /
1433
Temperature data
(without humidity)
Temperature range
°C /
°F
-70 to +180 /
-94 to 356
Temperature fluctuation
± K
0.1 to 0.5
0.1 to 0.5
0.1 to 0.5
Temperature uniformity (variation)
± K
0.1 to 1.3
0.2 to 1.8
0.3 to 2.0
Average heating up time acc. to IEC 60068-3-5
K/min.
5.5
5.0
4.8
Average cooling down time acc. to IEC 60068-3-5
K/min.
4.2
4.2
4.0
Max. heat compensation at 25 °C
/ 77°F
W
1500
3000
5000
Climatic data
(with humidity)
Temperature range
°C /
°F
+10 to +95 / 50 to 203
Temperature range
with optional compressed air dryer
°C /
°F
0 to +95 /
32 to 203
Temperature fluctuation
± K
0.1 to 1.0
0.1 to 1.5
0.1 to 1.0
Humidity range
% r.H.
10 to 98
10 to 98
10 to 98
Humidity range with optional compressed air dryer
% r.H.
5 to 98
5 to 98
5 to 98
Humidity fluctuation
± % r.H.
≤ 2.5
≤ 2.5
≤ 2.5
Содержание MKF 115
Страница 97: ...MKF MKFT E3 2 07 2017 page 97 115 20 10 Dimensions MKF 115 MKFT 115...
Страница 98: ...MKF MKFT E3 2 07 2017 page 98 115 20 11 Dimensions MKF 240...
Страница 99: ...MKF MKFT E3 2 07 2017 page 99 115 20 12 Dimensions MKFT 240...
Страница 100: ...MKF MKFT E3 2 07 2017 page 100 115 20 13 Dimensions MKF 720 MKFT 720...
Страница 102: ...MKF MKFT E3 2 07 2017 page 102 115...
Страница 103: ...MKF MKFT E3 2 07 2017 page 103 115...
Страница 104: ...MKF MKFT E3 2 07 2017 page 104 115 21 2 EU Declaration of Conformity for MKFT...
Страница 105: ...MKF MKFT E3 2 07 2017 page 105 115...
Страница 106: ...MKF MKFT E3 2 07 2017 page 106 115...
Страница 108: ...MKF MKFT E3 2 07 2017 page 108 115...
Страница 109: ...MKF MKFT E3 2 07 2017 page 109 115 22 Product registration...