MKF+ MKFT (E3.2) 07/2017
page 85/115
Fault description
Possible cause
Required measures
Condensation
Condensation at the samples.
Heating-up phase without bedew
protection.
Use the bedew protection
(chap. 10).
Heating up very fast.
Select lower heating up speed
(ramp).
Condensation or icing at the
sides of the inner chamber.
Set-point for a long time below ambi-
ent temperature, icing in the preheat-
ing chamber.
Defrost the chamber.
Condensation at the samples
or at the sides of the inner
chamber; notification “1H
PREHEAT PHASE” in the con-
troller display.
Rear power switch (20) turned on
less than 1 hour before operating the
chamber.
Turn on the rear power switch
(20) at least one hour before
operating the chamber.
Humidity
Humidity fluctuation:
Control accuracy of
±
2.5 %
r.H. is not reached.
Door gasket defective.
Replace door gasket.
Door opened very frequently.
Open doors less frequently.
Humidity fluctuation, together
with temperature fluctuation > 1
°C with a set-point approx. 3
°C above ambient temperature.
Place of installation too hot.
Select cooler place of installa-
tion or contact BINDER service.
No or low dehumidification.
Capillary tube blocked.
Contact BINDER service.
Not enough refrigerant.
No dehumidification; notifica-
tion “1H PREHEAT PHASE” in
the controller display.
Rear power switch (20) turned on
less than 1 hour before operating the
chamber.
Turn on the rear power switch
(20) at least one hour before
operating the chamber.
Icing at the sides of the inner
chamber.
Set-point was too long below ambient
temperature.
Defrost the chamber (chap. 15).
Condensation at the sides of
the inner chamber.
Combination of temperature/humidity
set-point values not in the optimum
range (see temperature humidity
diagram, Figure 19)
Select combination of tempera-
ture/humidity set-point values in
the optimum range (chap. 14).
Temperature set-point was too long
below ambient temperature, icing in
the preheating chamber.
Defrost the chamber (chap. 15)
Combination of temperature/humidity
set-point values leads to falling below
the dew point.
Select suitable combination of
temperature/humidity set-point
values.
Controller
No chamber function
(dark display).
Display mode “Standby” active.
Press any controller key.
Main power switch turned off.
Turn on the main power switch.
No entries to controller keypad
possible. Notification “KEY
LOCK” is displayed
Keyboard locking (option) activated.
Unlock keyboard locking (chap.
16.5).
No access to menu “User set-
tings”.
User code incorrect.
Contact BINDER service.
Wrong temperature alarms,
disturbance of temperature
accuracy
Temperature unit changed to °F.
Set temperature unit to °C
(chap. 6.4).
Chart recorder function: meas-
ured-value memory cleared,
information lost.
New setting of storage rate.
Change the storage rate ONLY
if the previously registered data
are no longer required (chap.
7).
Содержание MKF 115
Страница 97: ...MKF MKFT E3 2 07 2017 page 97 115 20 10 Dimensions MKF 115 MKFT 115...
Страница 98: ...MKF MKFT E3 2 07 2017 page 98 115 20 11 Dimensions MKF 240...
Страница 99: ...MKF MKFT E3 2 07 2017 page 99 115 20 12 Dimensions MKFT 240...
Страница 100: ...MKF MKFT E3 2 07 2017 page 100 115 20 13 Dimensions MKF 720 MKFT 720...
Страница 102: ...MKF MKFT E3 2 07 2017 page 102 115...
Страница 103: ...MKF MKFT E3 2 07 2017 page 103 115...
Страница 104: ...MKF MKFT E3 2 07 2017 page 104 115 21 2 EU Declaration of Conformity for MKFT...
Страница 105: ...MKF MKFT E3 2 07 2017 page 105 115...
Страница 106: ...MKF MKFT E3 2 07 2017 page 106 115...
Страница 108: ...MKF MKFT E3 2 07 2017 page 108 115...
Страница 109: ...MKF MKFT E3 2 07 2017 page 109 115 22 Product registration...