MKF+ MKFT (E3.2) 07/2017
page 31/115
CAUTION
Danger of overheating.
Damage to the voltage and frequency changer.
∅
Do NOT install the voltage and frequency changer in unventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
The voltage and frequency changer is equipped with
four castors. The rear castors can be easily locked
via the attached brakes
4.8
Electrical connection
4.8.1 Information on connecting the alternating climate chamber
The chambers are supplied ready for connection. They come with a fixed power connection cable of 2700
mm /
8.9 ft
in length and are equipped with3 internal overload releases against excess-current.
Model
Power plug
Nominal voltage
±
10% at the
indicated power frequency
Current
type
Chamber
fuse
MKF 115
MKFT 115
MKF 240
CEE plug 5-poles, 16 Amp
400 V at 50 Hz
3 N~
16 Amp
3 x internal
MKFT 240
CEE plug 5-poles, 16 Amp
400 V at 50 Hz
3 N~
16 Amp
3 x internal
MKF 720
MKFT 720
CEE plug 5-poles, 32 Amp
400 V at 50 Hz
3 N~
25 Amp
3 x internal
•
The domestic socket must also provide a protective conductor. Make sure that the connection of the
protective conductor of the domestic installations to the chamber’s protective conductor meets the lat-
est technology. The protective conductors of the socket and plug must be compatible!
•
Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the chamber’s type plate (left chamber side, bottom right-hand, chap. 1.4)
•
When connecting, please observe the regulations specified by the local electricity supply company and
as well as the VDE directives (for Germany). We recommend the use of a residual current circuit
breaker.
•
Pollution degree (acc. to IEC 61010-1): 2
•
Over-voltage category (acc. to IEC 61010-1): II
CAUTION
Danger of incorrect power supply voltage.
Damage to the equipment.
Check the power supply voltage before connection and start-up.
Compare the power supply voltage with the data indicated on the type plate.
See also electrical data (chap. 20.4).
To completely separate the chamber from the power supply, you must disconnect the power
plug. Install the chamber in a way that the power plug is easily accessible and can be easily
pulled in case of danger.
Содержание MKF 115
Страница 97: ...MKF MKFT E3 2 07 2017 page 97 115 20 10 Dimensions MKF 115 MKFT 115...
Страница 98: ...MKF MKFT E3 2 07 2017 page 98 115 20 11 Dimensions MKF 240...
Страница 99: ...MKF MKFT E3 2 07 2017 page 99 115 20 12 Dimensions MKFT 240...
Страница 100: ...MKF MKFT E3 2 07 2017 page 100 115 20 13 Dimensions MKF 720 MKFT 720...
Страница 102: ...MKF MKFT E3 2 07 2017 page 102 115...
Страница 103: ...MKF MKFT E3 2 07 2017 page 103 115...
Страница 104: ...MKF MKFT E3 2 07 2017 page 104 115 21 2 EU Declaration of Conformity for MKFT...
Страница 105: ...MKF MKFT E3 2 07 2017 page 105 115...
Страница 106: ...MKF MKFT E3 2 07 2017 page 106 115...
Страница 108: ...MKF MKFT E3 2 07 2017 page 108 115...
Страница 109: ...MKF MKFT E3 2 07 2017 page 109 115 22 Product registration...