MKF+ MKFT (E3.2) 07/2017
page 89/115
Chamber size
115
240
720
Climatic data
(with humidity)
Temperature range
°C /
°F
+10 to +95 /
50 to 203
Temperature range with optional compressed air
dryer
°C /
°F
0 to +95 /
32 to 203
Temperature fluctuation
± K
0.1 to 1.3
0.1 to 1.3
0.2 to 1.5
Humidity range
% r.H.
10 to 98
10 to 98
10 to 98
Humidity range with optional compressed air dryer
% r.H.
5 to 98
5 to 98
5 to 98
Humidity fluctuation
± % r.H.
≤ 2.5
≤ 2.5
≤ 2.5
Dew point temperature range
°C /
°F
+5 to +94 /
41 to 201
Dew point temperature range with optional com-
pressed air dryer
°C /
°F
-28 to +94 /
-18.4 to 201
Max. heat compensation at 25 °C /
77 °F
and 90 %
r.H. (without optional compressed air dryer)
W
400
400
1000
Electrical data
IP-system of protection acc. to EN 60529
IP
20
20
20
Nominal voltage (+/-10%) at 50 Hz power frequency
V
400
400
400
Current type
3N~
3N~
3N~
Nominal power
kW
4.80
6.80
11.00
Power plug: CEE plug 5-poles
Amp
16
16
32
Over-voltage category acc.
to
IEC 61010-1
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
Over-current release category B,
3 x internal
Amp
16
16
25
Electrical data of the voltage and frequency changer
IP-system of protection acc. to EN 60529
IP
23
23
23
Nominal voltage (+/-10%) 60 Hz (input side)
V
480 3N~
480 3N~
480 3N~
Nominal power
kW
9
9
13
Over-voltage category acc.
to
IEC 61010-1
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
Fuse
A
16
16
25
Environment-specific data
Noise level (mean value)
dB(A)
62
65
65
Noise level with optional compressed air dryer
(short-term) (mean value)
dB(A)
85
85
85
Noise level with voltage and frequency changer
(mean value)
dB(A)
67
67
67
Energy consumption
at +25 °C
/ 77 °F
and 60 % r.H.
Wh/h
1250
1500
3900
Filling weight of refrigerant R 404A (GWP 3750)
kg /
lbs.
2.00 /
4.41
2.20 /
4.85
5.00 /
11.02
Note:
Chambers with voltage and frequency changer: Average heating up time reduced by 0.3 K/min
each.
All technical data is specified for unloaded chambers with standard equipment at an ambient temperature
of +22 °C
±
3 °C /
71.6 °F ± 5.4 °F
and a power supply voltage fluctuation of +/-10%. Technical data is
determined in accordance to BINDER Factory Standard Part 1:2015 following DIN 12880:2007.
All indications are average values, typical for chambers produced in series. We reserve the right
to change technical specifications at any time.
If the chamber is fully loaded, the specified heating up and cooling down times may vary ac-
cording to the load.
Bringing a source of humidity into the inner chamber will affect the minimum humidity specifi-
cation and may affect the humidity accuracy.
Содержание MKF 115
Страница 97: ...MKF MKFT E3 2 07 2017 page 97 115 20 10 Dimensions MKF 115 MKFT 115...
Страница 98: ...MKF MKFT E3 2 07 2017 page 98 115 20 11 Dimensions MKF 240...
Страница 99: ...MKF MKFT E3 2 07 2017 page 99 115 20 12 Dimensions MKFT 240...
Страница 100: ...MKF MKFT E3 2 07 2017 page 100 115 20 13 Dimensions MKF 720 MKFT 720...
Страница 102: ...MKF MKFT E3 2 07 2017 page 102 115...
Страница 103: ...MKF MKFT E3 2 07 2017 page 103 115...
Страница 104: ...MKF MKFT E3 2 07 2017 page 104 115 21 2 EU Declaration of Conformity for MKFT...
Страница 105: ...MKF MKFT E3 2 07 2017 page 105 115...
Страница 106: ...MKF MKFT E3 2 07 2017 page 106 115...
Страница 108: ...MKF MKFT E3 2 07 2017 page 108 115...
Страница 109: ...MKF MKFT E3 2 07 2017 page 109 115 22 Product registration...