4.2.1 Automatic fresh water supply for humidifying system via water pipe
An enclosure inside the chamber contains the connection kit for water supply and wastewater. Install the
water supply connection using either the enclosed water hose or another pressure-resistant one. To ac-
complish this, remove the cover of the freshwater connection “IN” (18) (Figure 8) on the rear of the cham-
ber. Protect both ends of the hose with two of the four supplied hose clamps. Before turning on the
chamber, check the connection for leaks. Water supply is automatically effected via the freshwater con-
nection “IN” (18).
As the chamber only lets in water when required, there is no continuous water flow.
•
Supply pressure 1 to 10 bar when connecting to a water pipe.
•
Water type: deionized (demineralized) water
•
Water intake temperature NOT below +5 °C /
41 °F
and not exceeding 40 °C /
104 °F
.
•
The water intake shall be provided with a shut-off slide or water-tap.
•
For the water supply, fix the delivered adapter with hose olive on the thread at the rear of
the chamber.
•
Protect the water supply at one side with the supplied hose clamp.
4.2.2 Manual fresh water supply for humidifying system via internal freshwater can
If no house water connection with suitable water is available, you can manually supply water by filling the
freshwater can (total volume: 19 liters /
0.67 cu.ft.
up to the maximum level mark), which is located behind
the right door of the humidity generation module.
The cover of the water inlet valve must be screwed on the freshwater connection “IN” (18). Open the door
(F) (Figure 3) to access the filler neck of one of the water can. You cannot totally take out the water can
because of its fix connections. Fill the water can only up to ¾, up to the maximum level mark. When filling
it too much with the chamber turned on, the alarm message “
WATER LEVEL TOO HIGH
” is displayed on the
controller (chap. 13.1). Manually suck off the water, or operate the chamber with high temperature and
humidity values until the excess water is consumed. When filling it too much with the chamber turned off,
water can escape from the chamber. Thus, ensure not to fill the can by more than the maximum level
mark.
To guarantee humidification during 24 hours even at high humidity set-points with manual
water supply, we recommend filling the freshwater can (option) daily at the end of the day.
Содержание MKF 115
Страница 97: ...MKF MKFT E3 2 07 2017 page 97 115 20 10 Dimensions MKF 115 MKFT 115...
Страница 98: ...MKF MKFT E3 2 07 2017 page 98 115 20 11 Dimensions MKF 240...
Страница 99: ...MKF MKFT E3 2 07 2017 page 99 115 20 12 Dimensions MKFT 240...
Страница 100: ...MKF MKFT E3 2 07 2017 page 100 115 20 13 Dimensions MKF 720 MKFT 720...
Страница 102: ...MKF MKFT E3 2 07 2017 page 102 115...
Страница 103: ...MKF MKFT E3 2 07 2017 page 103 115...
Страница 104: ...MKF MKFT E3 2 07 2017 page 104 115 21 2 EU Declaration of Conformity for MKFT...
Страница 105: ...MKF MKFT E3 2 07 2017 page 105 115...
Страница 106: ...MKF MKFT E3 2 07 2017 page 106 115...
Страница 108: ...MKF MKFT E3 2 07 2017 page 108 115...
Страница 109: ...MKF MKFT E3 2 07 2017 page 109 115 22 Product registration...