5.6 Save/Print Test Results
Chapter 5 Run Test Operation
pH SAFE Reader Operator Manual - Doc. No. 120001-09
27
Options/System Settings/Erase Memory
function to erase all results
currently stored in the instrument memory (refer to Section 6.5.6,
Erase Memory).
5.6.2
Print pH
Use this option to output the test result to a Printer (or other receiving device,
see
Note
: below) connected to the RS-232 Serial port of the instrument. The
result is not stored
in the instrument memory.
When the
<Enter>
key is pressed to select use of this option, the screen updates
to briefly display the following message:
The data output from the serial port will be in the format illustrated in the
following example:
Figure 5-2
pH Result Printout
When the serial output has completed the screen briefly displays the following
message:
After displaying this message for approximately two seconds, the screen returns
to the
Run/Options
prompt.
NOTE:
The RS-232 Serial port hardware and corresponding firmware of the
BCSI pH1000
are designed to operate with a CBM Citizen model
iDP-3110 serial Printer. Alternatively, the instrument’s RS-232 Serial
port may be connected to a COM Port of an external computer using
a suitable RS-232 cable. The results output produced by the
BCSI
pH1000
can then be received using the Windows
®
HyperTerminal
application, or other serial data communication application. Refer to
Appendix A for complete information about serial communication
configuration requirements for this usage.
SELECT OPTION:
2. PRINT pH
OUTPUTTING
TEST RECORD
Blood Cell Storage Inc.
pH 1000 S/N PH04020731
Version X.X
REC#: 8
I.D.: ST13578A004MT
USER: USER1
DATE: 15 Mar 2020
TIME: 10:21
pH: 6.34
Instrument S/N
Software Ver. #
Record Sequence #
Platelet Storage
Bag ID
User ID
Test Date/Time
Result
OUTPUT COMPLETED