5.5 Read pH
Chapter 5 Run Test Operation
24
pH SAFE Reader Operator Manual - Doc. No. 120001-09
If a pHID has not been programmed into the instrument a warning will be
displayed. Please contact technical support to enable the appropriate pHID.
Next, the screen will advance to display the
Read pH
prompt. Proceed to Section
5.5, Read pH.
5.5
Read pH
Following the entry of the necessary identification information for the pH
measurement test of a platelet storage bag, the screen will display the following
prompt:
Proceed as follows:
3. Refer to Figure 5-1. Place the platelet storage bag onto the instrument
measurement platform and insert the Fiber Optic Probe of the instrument
into the opening at the top of the pH SAFE Sensor of the bag.
Figure 5-1
Insert Fiber Optic Probe into pH SAFE Sensor
4. Slide the pH SAFE Sensor of the platelet storage bag onto the pH SAFE
Reader Fiber Optic Probe until it snaps into place at the top of the sensor.
Ensure sensor is aligned with probe on reader. Avoid scraping the sensor on
the probe. When the test is successfully completed the screen updates to
display the platelet storage bag ID and the calculated pH test result, for
example:
5. From this screen, press any key to continue and display the Save/Print/
Discard Result menu. Proceed to Section 5.6, Save/Print Test Results.
READ pH: <Enter>
ABORT:
<Options>
Fiber Optic Probe
pH SAFE Sensor
ID: ST13578A004MT
pH: 6.87
Содержание pH SAFE PH1000