ACTAS set-up
1-7
5975636 ACTAS 2 170297
3. For nominal 120VAC operation, proceed as follows:
u
Change the line fuse to a 160mA slow blow type
fuse;
u
Remove the 230V soldered link on the module
(see fig. a);
u
Solder two links (marked as 110V) provided for
120V operation (see fig. b);
u
Remove the existing plug from the power cord
and strip the wires to attach the appropriate power
plug, ANSI 73.11.
Power cord with ANSI 73.11 plug
The wires of the power cord are colored in accord-
ance with the following code :
Green-and-yellow : Earth (safety earth)
White
: Neutral
Black
: Line (Live)
ANSI 73.11 plug
2. For nominal 230VAC operation, proceed as follows:
u
Change the line fuse to a 80mA slow blow type
fuse;
u
Remove the two 120V(marked as 110V) soldered
links on the module (see fig. b);
u
Solder one link provided for the 230V operation
(fig. a);
u
Remove the existing plug from the power cord
and strip the wires to attach the appropriate CEE7
power plug.
line fuse
T80mA
230 V link
230VAC operation
Fig. a
CEE 7 plug
110 V link
110 V link
Power cord with CEE 7 plug
The wires of the power cord are colored in
accordance with the following code :
Green-and-yellow : Earth (safety earth)
Blue
: Neutral
Brown
: Line (Live)
T80mA
230 V link
110 V link
110 V link
Fig. b
120VAC operation
line fuse
T160mA