5.10. Factory Gas Type Configuration
Gases listed as “factory gas types” are used less frequently than the more common basic gas types listed in Gas Type Menus
and Selection on page 47. Though used less frequently, the large list of supported factory gas types
are
available as H25-IR
PRO configuration options, but in a special Factory menu (versus the standard Gas Type menu) to minimize the sheer volume
of options in the Gas Type menu.
NOTE:
The Factory menu has restricted access and is not visible during normal operations. The procedure below
details how qualified operators may access the Factory menu. To return the Factory menu to restricted access
status, simply cycle the power to the H25-IR PRO.
If you have chosen to use one of these factory gas types (refer to Gas Type Menus and Selection on page 47), the gas type
shown in the Gas Type menu will reflect the your gas choice from the Factory menu. Four NDIR sensors are used to support
the listed factory gas types. These four NDIR sensors support the following factory gas types:
•
R600A and R290.
•
CO
2
.
•
SF
6
.
IMPORTANT:
Configuring the H25-IR PRO for a sample factory gas type requires (a) an appropriate NDIR sensor
installed in the instrument, and (b) gas type configuration in the Factory menu. Be sure you ordered an H25-IR
PRO with an appropriate NDIR sensor for the corresponding gas you wish to detect.
IMPORTANT:
Though setting your factory gas type is very straightforward, it
does
involve accessing the Factory
menu. Arbitrarily changing settings in the Factory menu or deviating from the procedure listed below may
produce non-standard operation, cause system failures, and void your factory warranty. Use caution when
making changes to the Factory menu.
Table 5-38. Configuring the H25-IR PRO to use a Factory Gas Type
Step
Button
Description
Sample Screen(s)
1
+
Enable the factory menu. With
the instrument OFF, press and
hold the ENT button while turning
ON the instrument.
You will hear a long beep, then the
Firmware Version screen is
displayed for approximately 15
seconds.
FIRMWARE
VERSION V1.09
Содержание H25-IR PRO
Страница 6: ...H25 IR PRO Instruction Manual vi P N 3015 5678 Rev 6 ...
Страница 8: ...H25 IR PRO Instruction Manual viii P N 3015 5678 Rev 6 ...
Страница 17: ...H25 IR PRO Instruction Manual Introduction P N 3015 5678 Rev 6 7 Figure 1 2 H25 IR PRO s ALQä Technology ...
Страница 39: ...H25 IR PRO Instruction Manual Menu System Overview P N 3015 5678 Rev 6 29 ...
Страница 109: ...H25 IR PRO Instruction Manual Maintenance P N 3015 5678 Rev 6 99 ...
Страница 120: ...H25 IR PRO Instruction Manual Declaration of Conformity P N 3015 5678 Rev 6 110 ...