Maintenance
H25-IR PRO Instruction Manual
96
P/N: 3015-5678 Rev 6
6.8. Firmware Upgrade Procedure
As updates to the instrument’s firmware become available, the instrument can be upgraded by using the procedure below.
NOTE:
The firmware upgrade procedure is sometimes referred to as
flashing memory
.
Table 6-13. Items Required for Upgrading Firmware
Item
Description
1
9-pin-Female to 9-pin-Male Serial Cable (P/N 104-4027)
2
Flash Utility Software and Latest Firmware Image File (downloaded from Website
http://www.mybacharach.com/downloads.htm)
3
Personal computer running Windows
®
95/98/2000/NT/XP/Vista/7
4
An open COM port on the computer
NOTE:
USB-to-Serial adapters
may
work, but are unsupported.
Table 6-14. Firmware Upgrade Procedure
Step
Description
1
Connect one end of a “9-pin female to 9-pin male" serial cable to the instrument’s RS-232 connector located at
the back of the instrument.
2
Connect the other end of the cable to an open COM port connector on the computer (it may be necessary to use
a 9-pin to 25-pin adapter to make the computer connection).
NOTE:
COM1 is the default port used by the Flash Memory Program. If COM1 is already in use by
another device, then connect the instrument to the next available COM port. The software will
need to be reconfigured as described in Step 8 to use this port.
3
Press and hold the
i
button on the front keypad and Turn ON the instrument. After the unit beeps quickly
several times release the
i
button. Pressing the ENT key will place the instrument in firmware upgrade mode.
NOTE:
The following steps assume that the operator is familiar with the Windows Operating
System, and is knowledgeable in creating folders, copying files, and navigating the file system using
Windows Explorer. If necessary, refer to the Windows help files for information on performing
these operations.
4
Create a folder on the computer’s hard drive named “C:\RFU" (Rabbit Field Utility).
5
Copy the flash utility software and the latest firmware image file (
xxx.bin
– where “xxx" is the name and version
number of the file) into the folder created in Step 4.
Содержание H25-IR PRO
Страница 6: ...H25 IR PRO Instruction Manual vi P N 3015 5678 Rev 6 ...
Страница 8: ...H25 IR PRO Instruction Manual viii P N 3015 5678 Rev 6 ...
Страница 17: ...H25 IR PRO Instruction Manual Introduction P N 3015 5678 Rev 6 7 Figure 1 2 H25 IR PRO s ALQä Technology ...
Страница 39: ...H25 IR PRO Instruction Manual Menu System Overview P N 3015 5678 Rev 6 29 ...
Страница 109: ...H25 IR PRO Instruction Manual Maintenance P N 3015 5678 Rev 6 99 ...
Страница 120: ...H25 IR PRO Instruction Manual Declaration of Conformity P N 3015 5678 Rev 6 110 ...