Operation
H25-IR PRO Instruction Manual
66
P/N: 3015-5678 Rev 6
Step
Button
Description
Sample Screen(s)
5
Press the ENT button to save
the new date and time and
return to the previous User
Setup screen.
NOTE:
The Date and Time are used in the Log Point data of the Measurement Log.
5.6.5. Log Limit
The Log Limit value is used in totalization applications and in conjunction with the relay outputs. Every time a measurement
is logged, its value is accumulated (until the Measurement Log is reset). Relay outputs can be energized if this accumulated
value ever exceeds a programmed log limit. The Log Limit is that setpoint.
NOTE:
For additional information on logging data and totalization applications, refer to Figure 5-8 on page 43
and Table 5-5 on page 44.
NOTE:
For relay configuration information, refer to the AT LOG LIMIT option in Table 5-21 (on page 63).
Table 5-24. Setting the Log Limit
Step
Button
Description
Sample Screen(s)
1
See Table 5-14
on page 53.
Follow the steps in Table 5-14
on page 53 to highlight the
Log Limit option.
SET DATE & TIME
qp
TO CHANGE
12/09/11 09:
30
t
u
TO MOVE
± SAVE
ESC=QUIT
<<
ADVANCED SETUP
>>
USER SETUPS 1-4
RELAY SETTINGS
MEASUREMENT LOG
DATE/TIME SET
LOG LIMIT
¯
qp
SELECT
± EDIT
Содержание H25-IR PRO
Страница 6: ...H25 IR PRO Instruction Manual vi P N 3015 5678 Rev 6 ...
Страница 8: ...H25 IR PRO Instruction Manual viii P N 3015 5678 Rev 6 ...
Страница 17: ...H25 IR PRO Instruction Manual Introduction P N 3015 5678 Rev 6 7 Figure 1 2 H25 IR PRO s ALQä Technology ...
Страница 39: ...H25 IR PRO Instruction Manual Menu System Overview P N 3015 5678 Rev 6 29 ...
Страница 109: ...H25 IR PRO Instruction Manual Maintenance P N 3015 5678 Rev 6 99 ...
Страница 120: ...H25 IR PRO Instruction Manual Declaration of Conformity P N 3015 5678 Rev 6 110 ...