H25-IR PRO Instruction Manual
Maintenance
P/N: 3015-5678 Rev 6
97
Step
Description
6
From the C:\RFU folder, run the flash memory program by double
clicking the file
rfu.exe
.
7
Select
Setup > Communications
.
8
Under “Serial Options”, select
Use Serial Connection.
Then set the
Baud Rate to
115200
, and select
Enable Processor Detection
.
NOTE:
COM1 is the default COM port. If necessary,
change the COM port to match the port that the
instrument was connected to in Step 1.
Click the OK button.
9
Select
Setup > File Locations
and verify that the coldload.bin, pilot.bin, and flash.ini files are all located in the
C:\RFU folder.
10
Click
OK
if these files are in the
correct folder. If necessary, use the
browse buttons in the Choose File
Locations dialog box to locate each
of these files.
11
Drag and drop the firmware file onto the program window. A plus sign (+) indicates the file is ready to copy.
Release the mouse and drop the firmware file onto the program window.