H25-IR PRO Instruction Manual
Maintenance
P/N: 3015-5678 Rev 6
95
Step
Description
20
Turn OFF instrument and remove the AC power cord.
NOTE:
Turning off the instrument returns the Factory menu to restricted access.
21
Once the unit is reassembled, reconnect the AC power cord and turn it back on. Allow it to complete the warm
up (60 seconds) and confirm no faults are displayed. This completes replacing the NDIR Sensor.
6.6. NDIR Sensor Adjustment
It may become necessary to adjust the emitter and detector components of the NDIR sensor without actually replacing the
NDIR sensor. Such instances may present themselves as clipping faults (fault code 8000), DigiPot range errors (fault code
0200), and other faults. If you suspect a sensor adjustment may be in order, or if you are directed to adjust your NDIR sensor’s
DigiPot settings by a Bacharach service representative, then use the
IR Emitter Adjustment
and
IR Detector Adjustment
components of the NDIR Sensor Replacement procedure found in the previous section.
NOTE:
“DigiPot” is a shortened version of
digital potentiometer
. The H25-IR PRO uses DigiPots to simplify certain
configuration processes by integrating the manual tweaking process of an internal potentiometer into the user
interface. If, for example, minor adjustments need to be made to the NDIR sensor power settings, the operator
can use the front panel buttons and display panel to make the adjustments (versus opening the case, attaching
a voltmeter to test points on the printed circuit board, and tweaking a potentiometer with a small screwdriver).
6.7. Spare Parts and Accessories
Table 6-12. Spare Parts and Accessories
Part Number
Description
3015-5668
O-Ring Kit, H25IR-Pro
3015-5818
External Filter Kit, H25IR-Pro (Includes 5 coalescing filters)
3015-5135
Probe Tip Filters (Qty 50)
3015-5099
Replacement Probe Tip
3015-5137
Replacement 8" Flexible Probe Gooseneck Assembly
3015-4670
Bezel, Case
4998-8986
Power Cord, USA (Type B)
0304-3462
Power Cord, India (Type D)
3015-5638
Power Cord (China)
0204-0020
Battery, Clock Backup
0604-2611
Fuse, 1A, Slow Blow (each)
See Table 6-10 on page 92
IR Sensors
Содержание H25-IR PRO
Страница 6: ...H25 IR PRO Instruction Manual vi P N 3015 5678 Rev 6 ...
Страница 8: ...H25 IR PRO Instruction Manual viii P N 3015 5678 Rev 6 ...
Страница 17: ...H25 IR PRO Instruction Manual Introduction P N 3015 5678 Rev 6 7 Figure 1 2 H25 IR PRO s ALQä Technology ...
Страница 39: ...H25 IR PRO Instruction Manual Menu System Overview P N 3015 5678 Rev 6 29 ...
Страница 109: ...H25 IR PRO Instruction Manual Maintenance P N 3015 5678 Rev 6 99 ...
Страница 120: ...H25 IR PRO Instruction Manual Declaration of Conformity P N 3015 5678 Rev 6 110 ...