H25-IR PRO Instruction Manual
Operation
P/N: 3015-5678 Rev 6
43
5.4.2. Log (Esc) Button
While in Measure Mode, the LOG (ESC) button adds the current non-zero gas measurement to the next
data storage location of the measurement log. The measurement log contains 50 data locations in which
leak rate data may be stored. Refer to Figure 5-8 and Table 5-5 on page 44.
NOTE:
The H25-IR PRO must be in Measure Mode and must be sampling a non-zero leak
rate in order to log leak rate data to the measurement log.
Figure 5-8. Graphical Representation of the Measurement Log
NOTE:
Use the Log Limit setting in conjunction with the programmable relay control settings to configure your
H25-IR PRO for totalizer applications. For example, consider a repetitive compressor test function at an air
conditioner manufacturer for inspecting up to 50 solder joints on each production unit for gas leaks. Consider a
corporate or environmental restriction that limits
each unit
to a leak rate of not more than 20 grams/yr. The
H25-IR PRO can be configured to measure all 50 points, quantify each leak rate, calculate the total of all 50 leak
rates, and activate an alert device (light, buzzer, alarm) via one or more of the output relays based on a
programmable Log Limit value. The measurement log can then be reset and the test repeated for the next device.
Содержание H25-IR PRO
Страница 6: ...H25 IR PRO Instruction Manual vi P N 3015 5678 Rev 6 ...
Страница 8: ...H25 IR PRO Instruction Manual viii P N 3015 5678 Rev 6 ...
Страница 17: ...H25 IR PRO Instruction Manual Introduction P N 3015 5678 Rev 6 7 Figure 1 2 H25 IR PRO s ALQä Technology ...
Страница 39: ...H25 IR PRO Instruction Manual Menu System Overview P N 3015 5678 Rev 6 29 ...
Страница 109: ...H25 IR PRO Instruction Manual Maintenance P N 3015 5678 Rev 6 99 ...
Страница 120: ...H25 IR PRO Instruction Manual Declaration of Conformity P N 3015 5678 Rev 6 110 ...