H25-IR PRO Instruction Manual
Operation
P/N: 3015-5678 Rev 6
77
NOTE:
A fault code of 0000 indicates that there are no current faults.
5.7.1. Current Faults Screen
Table 5-33. Accessing the Current Faults Screen
Step
Button
Description
Sample Screen
1
See Table 5-32
on page 75.
Follow the steps in Table 5-32
on page 75 to access the
Current Faults.
2
Press the ENT button to view
details of the current faults.
3
Press the Right Arrow button
to view the next fault (if one
exists).
Use the Left Arrow button to
return to the previous fault.
NOTE:
The Info (
i
) button can also be used to access the Diagnostics Menu and the CURRENT FAULTS screen.
NOTE:
The Info symbol (
i
) may also be displayed at the banner line on the display when a maintenance
condition warranting the users attention occurs.
NOTE:
The 4-digit fault code is an additive code that can represent multiple fault conditions. The first detected
fault is shown on the Search/Measure screen. The Diagnostics screen shows the fault “total”.
<<
DIAGNOSTICS
CURRENT FAULTS
0810
SENSOR DATA
qp
SELECT
± EDIT
FAULT 0800 >>
LOW AIR FLOW.
CHECK TIP FILTER
AND REPLACE, CHECK
HOSES FOR BLOCKAGE
<<
FAULT 0010
4-20MA CURRENT LOOP
IS OPEN OR CIRCUIT
HAS HIGH RESISTANCE.
CHECK JUMPER OR WIRES
ON REAR OF UNIT
Содержание H25-IR PRO
Страница 6: ...H25 IR PRO Instruction Manual vi P N 3015 5678 Rev 6 ...
Страница 8: ...H25 IR PRO Instruction Manual viii P N 3015 5678 Rev 6 ...
Страница 17: ...H25 IR PRO Instruction Manual Introduction P N 3015 5678 Rev 6 7 Figure 1 2 H25 IR PRO s ALQä Technology ...
Страница 39: ...H25 IR PRO Instruction Manual Menu System Overview P N 3015 5678 Rev 6 29 ...
Страница 109: ...H25 IR PRO Instruction Manual Maintenance P N 3015 5678 Rev 6 99 ...
Страница 120: ...H25 IR PRO Instruction Manual Declaration of Conformity P N 3015 5678 Rev 6 110 ...