21
RUN TEST MODE (MODELS 82X6 AND 82X7)
DUT POWER
Voltage
0
– 277 VAC Single Phase Unbalanced
(One Hot or Line conductor and One Neutral)
Current
16AAC max continuous
Voltage Display
Range:
0.0
– 277.0 VAC Full Scale
Resolution:
0.1 V
Accuracy:
(1.5
of reading
+
0.2V), 30.0
– 277.0VAC
Short Circuit Protection
23 AAC, Response Time < 3s
DELAY AND DWELL TIMER SETTINGS
Delay time setting
Range:
0.2
– 999.9 seconds
Resolution: 0.1 second
Accuracy:
(0.1% + 0.05 sec)
Dwell time setting
Range:
0.1
– 999.9 seconds (0 = Continuous)
Resolution: 0.1 second
Accuracy:
(0.1% + 0.05 sec)
TRIP POINT SETTINGS
Voltage:
Volt-Hi
Volt-LO
Range:
30.0
– 277.0 VAC
Resolution: 0.1 V
Accuracy:
(1.5
of setting
+
0.2 V), 30.0
– 277VAC
Current:
Amp-HI
Amp-LO
Range:
0.0
– 16.00 AAC
Resolution: 0.01 A
Accuracy:
(2.0
of setting
+
2 Counts)
Watts:
Power-HI
Power-LO
Range:
0
– 4500 W
Resolution: 1 W
Accuracy:
(5.0
of setting
+
3 Counts)
Power Factor:
PF-HI
PF-LO
Range:
0.000
– 1.000
Resolution: 0.001
Accuracy:
(8% of s 2 Counts)
Leakage Current:
Leak-HI
Leak-LO
Range:
0.00
– 10.00 mA (0 = OFF)
Resolution: 0.01 mA
Accuracy:
(2% of s 2 Counts)
Leakage current measuring resistor MD=2K
1%
METERING
Voltmeter
Range:
0.0
– 277.0 VAC
Resolution: 0.1 V
Accuracy:
(1.5
of reading
+
2 Counts), 30.0
–
277VAC
Ammeter
Range:
0.0
– 16.00 AAC
Resolution: 0.01 A
Accuracy:
(2.0
of reading
+
2 Counts)
Wattmeter
Range:
0
– 4500 W
Resolution: 1 W
Accuracy:
(5
of reading
+
3 Counts)
METERING
Содержание OMNIA II 8204
Страница 8: ...2 This Page Intentionally Left Blank ...
Страница 87: ...81 conditions The different combinations are represented in the table below ...
Страница 91: ...85 ...
Страница 92: ...86 ...
Страница 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Страница 141: ...135 and added to with the ADD command This will avoid constantly writing to the non volatile memory ...
Страница 142: ...136 ...
Страница 156: ...150 appear on the display ...
Страница 169: ...163 Print Report Example ...
Страница 181: ...175 ...