![Associated Research OMNIA II 8204 Скачать руководство пользователя страница 244](http://html1.mh-extra.com/html/associated-research/omnia-ii-8204/omnia-ii-8204_operation-and-service-manual_2989483244.webp)
238
Range 4
0.0uA ~ 126.0uA, frequency 15Hz
- 1MHz
Resolution
1uA
Additional Error
DC: add ±(2% of r 0.2% of
range) to base accuracy
15Hz < f < 100KHz : add ±(2% of
r 0.2% of range) to base
accuracy
100KHz < f < 1MHZ : add ±(2% of
r 0.5% of range) to base
accuracy
Range 5
0.0uA ~ 504uA, frequency 15Hz
- 1MHz
Resolution
1uA
Additional Error
DC: add ±(2% of r 0.2% of
range) to base accuracy
15Hz < f < 100KHz : add ±(2% of
r 0.2% of range) to base
accuracy
100KHz < f < 1MHZ : add ±(2% of
r 0.5% of range) to base
accuracy
Range 6
0.0uA ~ 600uA, frequency DC
–
100KHz
Resolution
1uA
Additional Error
DC: add ±(2% of r 0.2% of
range) to base accuracy
15Hz < f < 100KHz : add ±(2% of
r 0.2% of range) to base
accuracy
DISPLAY AUTO RANGE
Touch Current Display (Peak) (AC
Only)
Range 1
0.0uA ~ 32.0uA, frequency DC, 15Hz
– 1MHz
Resolution
0.1uA
Accuracy
15Hz < f <1MHz: ±(10% of r
2uA)
Range 2
28.0uA ~ 130.0uA, frequency DC,
15Hz - 1MHz
Resolution
0.1uA
Accuracy
15Hz < f <1MHz: ±(10% of r
2uA)
Range 3
120.0uA ~ 525.0uA, frequency DC,
15Hz - 1MHz
Resolution
0.1uA
Accuracy
15Hz < f <1MHz: ±(10% of r
2uA)
Содержание OMNIA II 8204
Страница 8: ...2 This Page Intentionally Left Blank ...
Страница 87: ...81 conditions The different combinations are represented in the table below ...
Страница 91: ...85 ...
Страница 92: ...86 ...
Страница 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Страница 141: ...135 and added to with the ADD command This will avoid constantly writing to the non volatile memory ...
Страница 142: ...136 ...
Страница 156: ...150 appear on the display ...
Страница 169: ...163 Print Report Example ...
Страница 181: ...175 ...