116
CONT, RUN or LCT must be used. The parameters <
p1,p2>
etc. indicate the
individual settings for each test. All parameters must be included with the command
and should appear in the same order that is shown on the actual setting screens. The
correct order is shown in the tables below. When the scanner is installed it should be
the last parameter for all test types except GND. For the GND test type, the scanner
settings should be inserted between offset and frequency. The list of parameters can
also be found in the default parameters section of the manual.
The parameter values should use complete text and not use the coded values that are
associated with the individual parameter setting commands. Such as "ON" and "OFF"
and any toggle field that use words or phrases like "OPEN", "CLOSE". The LS?
companion command will also list all parameters in complete text as they appear on
the setting screen.
ACW
DCW
IR
1
Voltage
Voltage
Voltage
2
HI-Limit T
HI-Limit
HI-Limit
3
LO-Limit T
LO-Limit
LO-Limit
4
Ramp Up
Ramp Up
Ramp Up
5
Dwell Time
Dwell Time
Delay Time
6
Ramp Down
Ramp Down
Ramp Down
7
Arc Sense
Charge-LO
Charge-LO
8
HI-Limit Real
Arc Sense
DUT Output
1
9
LO-Limit Real
Ramp-HI
(ON/OFF)
Scanner Setup
2
10
Frequency
Arc Detect
(ON/OFF)
11
Arc Detect
(ON/OFF)
Continuity
(ON/OFF)
12
Continuity
(ON/OFF)
DUT Output
1
13
DUT Output
1
Scanner Setup
2
14
Scanner Setup
2
GND
CONT
RUN
LCT
1
Current
HI-Limit
Voltage-HI
Leakage-HI
2
Voltage
LO-Limit
Voltage-LO
Leakage-LO
3
HI-Limit
Dwell
Amp-HI
Voltage-HI
4
LO-Limit
Offset
Amp-LO
Voltage-LO
5
Dwell
Scanner
Channel
Dwell Time
Delay Time
6
Offset
Continuity Delay Time
Neutral
(OPEN/CLOSED)
7
Scanner
Channel
Leakage-HI
Reverse
(ON/OFF)
8
Frequency
Leakage-LO
Ground
(OPEN/CLOSED)
Содержание OMNIA II 8204
Страница 8: ...2 This Page Intentionally Left Blank ...
Страница 87: ...81 conditions The different combinations are represented in the table below ...
Страница 91: ...85 ...
Страница 92: ...86 ...
Страница 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Страница 141: ...135 and added to with the ADD command This will avoid constantly writing to the non volatile memory ...
Страница 142: ...136 ...
Страница 156: ...150 appear on the display ...
Страница 169: ...163 Print Report Example ...