INSULATION RESISTANCE TEST MODE
Voltage Setting Range:
30
– 1000 VDC
Resolution: 1 V
Accuracy:
(2% of s 2 counts)
Charging
Current
Maximum
>20mA peak
Voltage Display Range:
0
– 1000 V
Resolution: 1 V
Accuracy:
(2% of r 2 counts)
Resistance
Display
Range:
0.05M
– 50000 M
(4 Digit, Auto Ranging)
Resolution:
30
– 499 VDC
500
– 1000VDC
M
M
M
0.001
0.050
– 1.999
0.050
– 9.999
0.01
2.00
– 19.99
10.00
– 99.99
0.1
20.0
– 199.9
100.0
– 999.9
1
200
– 50000
1000
– 50000
Accuracy:
50
– 499V
0.05M
– 999.9M,
(7% of r2 counts)
500
– 1000V
0.05M
– 999.9M,
(2% of r2 counts)
1000M
– 9999M
(5% of r2 counts)
10000M
– 50000M,
(15% of r2
counts)
Charge-LO
Range:
0.000
– 3.500
A or Auto Set
HI and LO
–
Limit
Range:
0.05M
– 99.99M
Resolution: 0.01M
Range:
100.0M
– 999.9M
Resolution: 0.1M
Range:
1000M
– 50000M
Resolution: 1M
(HI
– Limit: 0 = OFF)
Accuracy: Same as Resistance Display Accuracy
Ramp Timer
Range:
Ramp-Up:
0.1
– 999.9 sec
Ramp-Down:
0.0,
1.0
–999.9 sec
Resolution: 0.1 sec
Accuracy:
(0.1% + 0.05 sec)
Delay Timer
Range:
0.5
– 999.9 sec (0 = Continuous)
Resolution: 0.1 sec
Accuracy:
(0.1% + 0.05 sec)
Ground Fault
Interrupt
GFI Trip Current:
0.4mA - 5.0mA
HV Shut Down Speed:
1mS
Содержание OMNIA II 8204
Страница 8: ...2 This Page Intentionally Left Blank ...
Страница 87: ...81 conditions The different combinations are represented in the table below ...
Страница 91: ...85 ...
Страница 92: ...86 ...
Страница 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Страница 141: ...135 and added to with the ADD command This will avoid constantly writing to the non volatile memory ...
Страница 142: ...136 ...
Страница 156: ...150 appear on the display ...
Страница 169: ...163 Print Report Example ...
Страница 181: ...175 ...