195
more information.
At the end of the Verification process, a pop-up message will appear indicating the
outcome of the process. If the instrument passes the Verification (test failure,
indicating the fail detectors are working properly) the RESET button will illuminate, the
alarm will sound and the following pop-up message will appear over the metering
screen:
If the instrument fails the Verification (test pass, indicating the fail detectors are not
working properly) the TEST button will illuminate and the following pop-up message
will appear over the metering screen: