123
COMMAND
NAME
TEST
TYPES
VALUE
EC <
value
>
EC?
Edit Current
GND
1.00 - 40.00A
ECH <
value
>
ECH?
Edit Current-HI
RUN
0.00 - 16.00A
ECL <
value
>
ECL?
Edit Current-Lo
RUN
0.00 - 16.00A
EDE <
value
>
EDE?
Edit Delay
IR
RUN
LCT
0.0 - 999.9s
EDO {1|0}
EDO?
Edit DUT-Output
Voltage
ACW
DCW
IR
1= On, 0=Off
EDW <
value
>
EDW?
Edit Dwell
ACW
DCW
GND
CONT
RUN
0.0 - 999.9s
EF {1|0}
EF?
Edit Frequency
ACW
GND
1=60Hz, 0=50Hz,
2=400Hz (option),
3=800Hz (option),
4=SYNC
EG {1|0}
EG?
Edit Ground
LCT
1=Open, 0=Close
EH <
value
>
EH?
Edit HI-Limit
DCW
IR
GND
CONT
0.0 - 20000uA
0.05 - 50000M
0 - 600m
0.00 - 10.00
EHR <
value
>
EHR?
Edit HI-Limit-R
ACW
0.000 - 50.00mA,
99.99mA (Models 825x)
EHT <
value
>
EHT?
Edit HI-Limit-T
ACW
0.000 - 50.00mA,
100.00mA (Models 825x)
EL <
value
>
EL?
Edit LO-Limit
DCW
IR
GND
CONT
0.0 - 20000uA
0.05 - 50000
0 - 600m
0.00 - 10000
ELR <
value
>
ELR?
Edit LO-Limit-R
ACW
0.000 - 50.00mA,
99.99mA (Models 825x)
ELT <
value
>
ELT?
Edit LO-Limit-T
ACW
0.000 - 50.00mA,
100.00mA (Models 825x)
ELM {1|0}
ELM?
Edit Leakage
Mode
LCT
1=Peak, 0=RMS
Содержание OMNIA II 8204
Страница 8: ...2 This Page Intentionally Left Blank ...
Страница 87: ...81 conditions The different combinations are represented in the table below ...
Страница 91: ...85 ...
Страница 92: ...86 ...
Страница 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Страница 141: ...135 and added to with the ADD command This will avoid constantly writing to the non volatile memory ...
Страница 142: ...136 ...
Страница 156: ...150 appear on the display ...
Страница 169: ...163 Print Report Example ...
Страница 181: ...175 ...