84
Measuring Device
The Measuring Device is an impedance circuit that simulates the human body. The
Measuring Device is specified differently for different applications and specifications.
Repeatedly pressing the “Measure Device” soft key will allow you to select one of the
many measuring devices listed in the table below. The table below indicates the
available measuring devices and the appropriate agency standard to which it
correlates.
Cross
Reference
Number
Agency Specification
Application
1
UL544 Non patient equipment
Medical Equipment
2
UL544 Patient care equipment
Medical Equipment
3
IEC601-1, UL2601, EN60601-1
Medical Equipment
4
UL1563
Electric Spas,
Equipment Assemblies and
Associated Equipment
5
IEC60990 Fig4-U2, IEC60950
Laboratory Equipment
Information Technology
6
IEC60990 Fig5-U3
Laboratory Equipment
Information Technology
External
User configurable
User configurable
Frequency
Check
The Leakage Current test can be
configured to verify the bandwidth
of the leakage current “voltmeter”,
by setting the Probe configuration,
“Probe-HI to Probe-LO”, and
selecting the measuring device
“Frequency Check”.
The selection is intended to be
used for meter verification only
and has no application for
normal testing.
Measuring Device Diagrams
Diagrams of the measuring devices are shown in the images below.
Содержание OMNIA II 8204
Страница 8: ...2 This Page Intentionally Left Blank ...
Страница 87: ...81 conditions The different combinations are represented in the table below ...
Страница 91: ...85 ...
Страница 92: ...86 ...
Страница 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Страница 141: ...135 and added to with the ADD command This will avoid constantly writing to the non volatile memory ...
Страница 142: ...136 ...
Страница 156: ...150 appear on the display ...
Страница 169: ...163 Print Report Example ...
Страница 181: ...175 ...