![Associated Research OMNIA II 8204 Скачать руководство пользователя страница 136](http://html1.mh-extra.com/html/associated-research/omnia-ii-8204/omnia-ii-8204_operation-and-service-manual_2989483136.webp)
130
Reads the instrument’s scanner port status. The query will return either a 0 (none), 1
(external 1), 2 (external 2) or 3 (both external 1 and 2) which describes the number of
local scanners connected to the instrument.
SS?
Reads the selected step from the instrument.
7.4.7. IEEE 488.2 Common Commands
These commands are required by the IEEE-488.2 standard with the exception of
*PSC, *PSC?. Most of these commands are not available over the RS-232 bus
except for the *IDN? command which can be used to retrieve the instrument
identification information, and the four status reporting commands *ESR?, *ESE,
*ESE? and *STB?.
COMMAND
NAME
DESCRIPTION
*IDN?
Identification Query
ARI, Model Number, Serial
Number, Firmware Revision
*RST
Reset Command
Resets OMNIA
*TST?
Self-Test Query
00H=OK
01H=TEST EEPROM ERROR
*CLS
Clear Status Command
Clear Standard Event Status
Register
Clear Service Request Register
*OPC
Operation Complete Command
When TEST command ok setting
ESR BIT0 =1
*OPC?
Operation Complete Query
1 = TEST completed ok
0 = TEST in process
*WAI
Wait-to-Continue Command
*PSC {1|0}
Power-on Status Clear Command 1 = Power-on clear enable
registers
0 = Power-on load previous
enable registers
*PSC?
Power-on Status Clear Query
*ESR?
Standard Event Status Register
Query
0 - 255
*ESE <
value
> Standard Event Status Enable
Command
value
= 0 - 255
*ESE?
Standard Event Status Enable
Query
0 - 255
*STB?
Read Status Byte Query
Read Status Byte
*SRE
<
value
>
Service Request Enable Command
value
= 0 - 255
*SRE?
Service Request Enable Query
0 - 255
*IDN?
Содержание OMNIA II 8204
Страница 8: ...2 This Page Intentionally Left Blank ...
Страница 87: ...81 conditions The different combinations are represented in the table below ...
Страница 91: ...85 ...
Страница 92: ...86 ...
Страница 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Страница 141: ...135 and added to with the ADD command This will avoid constantly writing to the non volatile memory ...
Страница 142: ...136 ...
Страница 156: ...150 appear on the display ...
Страница 169: ...163 Print Report Example ...
Страница 181: ...175 ...