87
8.9.6. IEEE 488.2 Common Commands
These commands are required by the IEEE-488.2 standard with the exception of *PSC, *PSC?.
Most of these commands are not available over the RS-232 bus except for the *IDN? command
which can be used to retrieve the instrument identification information, and the four status
reporting commands *ESR?, *ESE, *ESE? and *STB?
Command
Name
Description
*IDN?
Identification Query
Associated Research Inc., Model
Number, Serial Number, Firmware
Revision
*RST
Reset Command
Resets HYPOTULTRA III
*TST?
Self-Test Query
00H=OK
01H=TEST EEPROM ERROR
*CLS
Clear Status Command
Clear Standard Event Status Register
Clear Service Request Register
*OPC
Operation Complete Command
When all selected pending
operations complete, ESR BIT0=1
*OPC?
Operation Complete Query
When all selected pending
operations complete, Output Queue=1
*WAI
Wait-to-Continue Command
*PSC {1|0}
Power-on Status Clear Command
1 = Power-on clear enable registers
0 = Power-on load previous enable
registers
*PSC?
Power-on Status Clear Query
*ESR?
Standard Event Status Register Query 0 - 255
*ESE <
value
> Standard Event Status Enable
Command
value
= 0 - 255
*ESE?
Standard Event Status Enable Query
0 - 255
*STB?
Read Status Byte Query
Read Status Byte
*SRE <
value
> Service Request Enable Command
value
= 0 - 255
*SRE?
Service Request Enable Query
0 - 255
*IDN?
Read the instrument identification string. Company =Associated Research Inc.
*RST
Reset the instrument to original power on configuration. Does not clear Enable register for
Standard Summary Status or Standard Event Registers. Does not clear the output queue. Does
not clear the power-on-status-clear flag.
*TST?
Performs a self-test of the instrument data memory. Returns 0 if it is successful or 1 if the test
fails.
Содержание HYPOTULTRA III 7620
Страница 7: ......
Страница 8: ......
Страница 130: ...122 configure it through the drive s properties in Windows ...
Страница 142: ...134 This screen will appear after both steps 1 and 2 CALIBRATION Automatic calibration process ...