112
To navigate to the Data Card Settings menu, select the <more> soft key from the Perform
Tests screen. Then select the “DCard” soft key. A brief description of the settings
follows:
Results Limit
Highlight the Results Limit parameter using the “
∧
,
∨
” soft keys. When the Results
Limit parameter is highlighted, press the “Edit” soft key.
The Results Limit allows the user to receive an alert message on the display of the
HypotULTRA III when the total number of results saved on the Data Storage Card’
s
internal Flash memory has exceeded the predefined limit. Use the numeric keypad to set
this value from the default value of “0” (disables the function) to 100,000. Press the
ENTER key to accept the new setting or the EXIT key to cancel and return to the original
setting. At the conclusion of a test file, if the total results on the internal Flash memory
exceed the Results Limit, the following message will display:
D A T A C A R D A L E R T
R e s u l t s L i m i t h a s b e e n
e x c e e d e d . P r e s s E N T E R
t o s h o w t h i s m e s s a g e
a g a i n , E X I T t o c a n c e l .
The message gives two options: Pressing Enter will clear the message and display it again
at the end of the next test file execution. Pressing Exit will clear the message and not
display it again unless the unit is re-booted, the Results Limit parameter is increased, or
the user deletes all previous results and the total results are exceeded once more.
Time Limit
Highlight the Time Limit parameter using the “
∧
,
∨
” soft keys. When the Time Limit
R e s u l t s L i m i t 0 0 0 0 0 0
T i m e L i m i t 0 d a y s
^
R S 4 8 5 A d d r e s s 1 0
T e s t R e s u l t A L L
v
S e r i a l P o r t R S 2 3 2
B a r c o d e I n p u t S E R / P R O D
E d i t
A C W T o t a l S e t t i n g s
M 5 0 - 8 _
P
9 9 9 . 9 s 1 0 . 0 0 m A R
R e s u l t
S y s t e m
D C a r d
< t o p >