31
The Perform Tests screen will now be displayed. The Perform Test screen will appear as
follows:
A C W T o t a l S e t t i n g s M e m N a m e * *
M 5 0 - 8 _
P
9 9 9 . 9 s 1 0 . 0 0 m A R
M e m o r y
S t e p
E d i t
< m o r e >
A C W T o t a l S e t t i n g s
M 5 0 - 8 _
P
9 9 9 . 9 s 1 0 . 0 0 m A R
R e s u l t
S y s t e m
< t o p >
The HypotULTRA III comes with the Memories/Steps and System loaded with the factory
defaults. The factory defaults are as follows:
Test
Type Parameter
Value
ACW
Voltage
1240V
Max Lmt T
10.00mA
Min Lmt T
0.000mA
Max Lmt R
10.00mA
Min Lmt R
0.000mA
Display
Total
Ramp UP
0.1s
Dwell
1.0s
Ramp DN
0.0s
Connect
OFF
Frequency
60Hz
Arc Detect
OFF
Arc Sense
5
Continuity
OFF
Scanner
OOOOOOOO
Prompt
System Presets
Setup Sys.
PLC Remote
OFF
Single Step
OFF
Fail Stop
ON
Alarm
5
Contrast
5