93
channel set to the return point), H (scanner set to the high voltage point or continuity output)
and O (OFF). Once the channel is set, press ENTER to accept the new setting or EXIT to
cancel.
Reviewing the scanner settings
From the Perform Tests screen, press the “Step” soft key. The soft keys will now change to
Step -, Step +, and Edit. The display will appear as follows:
A C W T o t a l S e t t i n g s H L H L H O O O
M 5 0 - 8 _
P
9 9 9 . 9 s 1 0 . 0 0 m A R
S t e p –
S t e p +
E d i t
When the scanner is installed, a scanner status display will replace the memory name. This
feature allows you to quickly review the scanner settings of the individual steps by pressing the
“Step + and Step-” soft keys.
Running Scanner Status
When the scanner is installed, a live scanner status will appear in the upper right comer of the
display while a test is being performed.
A C W T o t a l D w e l l M e m N a m e
M 5 0 - 8 _
P
2 . 3 s 1 0 . 0 0 m A R
+
_
The Live Scanner Status gives direct visual feedback of the state of all eight of the scanner
channels. The numbers indicate the channel numbers from left to right and the blocks represent
channels 1-8. A darkened “H” block indicates that channel is set to be a High
Voltage/Continuity output. A darkened “L” block indicates the channel is set to be a return
connection. An open block indicates that the scanner channel is set to OFF. For the example
illustrated below the channels are set as follows: Channel 1= High Voltage, Channel 2= Return,
Channel 3= OFF, Channel 4 =High Voltage, Channel 5=OFF, Channel 6= Return, Channel 7 =
OFF, Channel 8= OFF.
Scanner settings for the
current step
Live Scanner
Status