96
In the OFF mode, the printer card and printer menuing are disabled.
Test Result
From the Print Format setting screen, press the “Test Result” soft key. As the soft key is
pressed, the display will toggle between the words “All” and “Fail only”. This command selects
which test results will be printed, “All”, selects printing all test results and “Fail only”, selects
printing only failed test results.
Test Setting
From the Print Format setting screen, you may turn the Test Setting print enable ON and OFF
by pressing the “Test Setting” soft key. This command selects printing the test parameters as
well as the test results.
Device ID
The Device ID number is a counter that increments once every time a test is performed. The
number can be set to match the exact serial number or some portion of actual serial number of
the item under test or just used as an identifier or tracking number. From the Print Format
setting screen, use the numeric keypad to select the initial Device ID. Finish by pressing the
ENTER key. This number is not saved as part of the non-volatile system parameters; therefore,
each time the power to the instrument is turned off the number will be reset to 1.
Form Feed
From the Print Format setting screen, you may turn the Form Feed ON and OFF by pressing the
“Form Feed” soft key. This command forces a form feed after each complete test sequence.
When the form feed is turned off, the signature line will not appear. When the form feed is
turned off the pages will break whenever the page is determined to be full by the printer.