63
5.5. Displayed Messages
Directly above the Metering screens is the Test status display. This portion of the display is
active during the test and allows you to view the type of test being performed and status of the
test step. At the end of a test, the Test status display will inform you either that the test has
passed or failed. The following is an example of the Test status display:
A C W T o t a l D w e l l
A C W T o t a l D w e l l
M 5 0 - 8 _
P
2 . 3 s 1 0 . 0 0 m A R
+
_
5.5.1. Test Status Messages
The following is a complete listing of the messages that appear in the Test status display and an
explanation for each.
Dwell
This message appears on the display while a test is in process, the values update in real time.
Delay
This message appears on the display while a test is in process, the values update in real time.
Ramp Up
This message appears on the display at the beginning of the test, when the output voltage is
ramping up.
Ramp Down
This message appears on the display at the end of the dwell cycle, when the output voltage is
ramping down.
Pass
This message appears on the display when the test process is complete and the DUT passed the
test.