The parameter values should use complete text and not use the coded values that are associated
with the individual parameter setting commands such as "ON" and "OFF" for toggle field or
words such as “Real” or “Total”. The LS? companion command will also list all parameters in
complete text in the order as they appear in the following table, preceded by the step number.
ACW
DCW
IR
CONT
1
Voltage
Voltage
Voltage
Max-Limit
2
Max-Lmt T
Max-Lmt
Max-Lmt
Min-Lmt
3
Min-Limit T
Min-Limit
Min-Limit
Dwell
4
Ramp Up
Ramp Up
Ramp Up
Offset
5
Dwell Time
Dwell Time
Delay Time
Connect
6
Ramp Down
Ramp Down
Ramp
Down Internal
Scanner (opt)
7
Arc Sense
Charge-LO
Charge-LO
External
Scanner
8
Max-Lmt Real
Arc Sense
Connect
9
Min-Lmt Real
Ramp-HI
(ON/OFF)
Internal
Scanner (opt)
10
Frequency
(50/60)
Arc Detect
(ON/OFF)
External
Scanner
11
Arc Detect
(ON/OFF)
Continuity
(ON/OFF)
12
Continuity
(ON/OFF)
Connect
13
Display
(Real/Total)
Internal Scanner (opt)
14
Connect
External Scanner
15
Internal Scanner (opt)
16
External Scanner
SP <prompt message>
Adds or edits a prompt message for the active step.
SP
Removes or deletes the prompt that had been created for the active step.
8.9.3. Test Parameter Editing Commands and Companion Queries
These commands are used to modify the test parameter within each step. These commands
require a parameter value to be included with the command. The companion query command
will read the parameter. The writing of the parameter requires that the unit not be included with
the value, only the numeric value should be included with the command. Also, when the query
commands are used the response will not include the unit’s characters. Many of the commands
will function the same way for multiple test types; however, the input range may be different and
therefore used a different possible set of values.