19
Return Connection
When the instrument Return is grounded, any internal and external stray leakage will be
monitored due to currents that flow from High Voltage to earth ground (such as from HV to the
chassis of the instrument). This current is inherent and will cause errors when trying to monitor
very low leakage currents in the microampere range.
It is important to note that grounding the return lead will disable the Smart GFI function. Smart
GFI allows the user to automatically configure the instruments return configuration. When the
Return lead is earth grounded, the GFI circuit is disabled and the instrument operates in a
grounded return mode of operation.
Please refer to section
4.1.2 System Parameters
for a complete explanation of the GFI
function.
2.2.4. Environmental Conditions
This equipment is intended for indoor use only. The equipment has been evaluated according to
Installation Category II and Pollution Degree 2 as specified in IEC 664.
This instrument may be operated in temperatures from 32° - 104° F (0° - 40° C).
Relative humidity of 0 to 80%.
Altitude up to 6,560 feet (2,000 meters).
Storage and Shipping Environment
This instrument may be stored or shipped in environments with the following limits:
Temperature.........................-40° - 167° F ( -40° - 75°C)
Altitude … … … … … … 50,000 feet (15,240 meters)
The instrument should also be protected against temperature extremes that may cause
condensation within the instrument.