11
Ambulatory Patient Transfer
1.
If in operation, press green power button to turn off therapy control unit. surface will begin to deflate.
2.
adjust height of both surfaces to comfortable levels for patient transfer.
3.
Lower side rails when applicable.
4.
Lock caster wheel brakes and caster wheel steering on both surfaces.
5.
assist patient in ambulatory transfer, following all applicable safety rules and institution protocols.
6.
Center patient on the first step select™ mrs surface.
7.
raise and lock side rails on both sides of bed, if decision has been made to use side rails (refer to the
Safety Information
and
Risks and Precautions
sections in the
Introduction
chapter of this manual).
8.
press green power button to activate therapy control unit and adjust air pressures. see the
Air Pressure Adjustment
section of this chapter.
9.
With air pressures adjusted, verify patient comfort.
10. Lower bed height to lowest level.
11. activate warmer and adjust temperature, if desired.
Non-ambulatory Patient Transfer
1.
press green power button on therapy control unit to activate blower.
2.
press the Instaflate™ Inflation function (If) to fully inflate cushions and create a firm surface for patient transfer.
3.
perform steps 2-7 from Ambulatory Patient Transfer section.
4.
Deactivate the Instaflate™ If and adjust air pressures. see Air Pressure Adjustment section of this chapter.
5.
perform steps 9-11 of Ambulatory Patient Transfer section.
Patient Transfer from the First Step Select
™
Unit
Ambulatory Patient Transfer
1.
press green power button to turn off therapy control unit. surface will begin to deflate.
2.
adjust height of bed to same level as surface to which patient is being transferred.
3.
Lock caster wheel brakes and caster wheel steering on both surfaces.
4.
Lower side rails when applicable.
5.
transfer patient following all applicable safety rules and institution protocols.
Non-ambulatory Patient Transfer
1.
press Instaflate™ If to fully inflate cushions and create a firm surface for patient transfer.
2.
perform steps 2-5 from
Ambulatory Patient Transfer
section.
Air Pressure Adjustment
Use Pressure Adjust button to adjust pressure in each individual section of surface. Air pressures are set by using either HT/WT
pre-set or manual method.
1.
press pressUre aDJUst button shown at right.
2.
Display shown at right will appear.
PRESSURE
HT/WT
ADJUST
MANUAL
EXIT
Содержание First Step Select
Страница 2: ......
Страница 4: ......
Страница 20: ...20 ...
Страница 24: ...24 ...
Страница 26: ...26 ...
Страница 36: ...www ArjoHuntleigh com Manufactured for ArjoHuntleigh AB Hans Michelsensgatan 10 211 20 Malmö Sweden ...
Страница 38: ......
Страница 40: ......
Страница 56: ...56 ...
Страница 60: ...60 ...
Страница 62: ...62 ...
Страница 72: ...www ArjoHuntleigh com Fabriqué pour ArjoHuntleigh AB Hans Michelsensgatan 10 211 20 Malmö Suède ...
Страница 73: ...Colchón médico antiescaras First Step Select Manual del usuario 215238 AH Rev D 02 2014 with people in mind ...
Страница 74: ......
Страница 76: ......
Страница 92: ...92 ...
Страница 96: ...96 ...
Страница 98: ...98 ...
Страница 108: ...www ArjoHuntleigh com Fabricado por ArjoHuntleigh AB Hans Michelsensgatan 10 211 20 Malmö Suecia ...