26
Instructions For Use
EN
NOTE. The TOCO-T transducer and the TOCO-E transducer cannot be used
simultaneously.
2.2.6 Connecting to the Monitor
Use the interface cable supplied to connect the FTS-3 system to the monitor. Switch on
the monitor and the FTS-3 system.
2.2.7 Adjusting the Working Channel
Wireless transmission can be interrupted by other transmitters operating on the same
frequency. If there are several FTS-3 units in one ward, all base stations must be set to
different channel numbers. The operating channel is set using the adjustment buttons
located on the rear panel of the base station. This is performed as follows:
1. Put all the transducers back in the docking slots.
2. Use the adjustment buttons on the rear panel of the base station to select the new
channel number
. The channel range is 1-14.
3. Restart the system when it enters the charging interface.
NOTE. The working channel number used by a system cannot be duplicated
with that used by a device of the same type.
2.3
Basic Operation
2.3.1 Charging the Transducers
Place the transducer in the docking slot and it displays the charging state on the
transducer screen.
NOTE. When the TOCO-E transducer is connected with the DECG or MECG
cable, it cannot be charged due to poor contact with the charging points.
Please remove the DECG or MECG cable before charging
CAUTION. Wait for 2 minutes to use the transducers after charging.
2.3.2 Charge the Battery
Pay attention to the battery level during the monitoring process. The battery symbol
is displayed in the top right corner of the transducer screen. A low battery level may
influence the monitoring.