25
Instructions For Use
EN
2.2.3 Installing the System
FTS-3 should be placed on a flat surface. It should be away from any device with
strong radiation and avoid being in a shielded room. More than 2 similar systems
should be kept at a distance of over 1.5m.
Alternatively, it can be installed on a wall or a trolley approved by Huntleigh. Consult
the sales representative for more information.
CAUTION. Installation must be carried out by qualified personnel authorised
by the manufacturer.
CAUTION. If you choose to install FTS-3 on a trolley, it is the user’s
responsibility to ensure their integrity and solidity evaluated by
a registered, professional structural or mechanical engineer and
compliance with all local regulations. The manufacturer will not be
responsible for the failure and loss of any improper installation.
2.2.4 Connecting the Power Cable
Make sure the AC power supply of the system complies with the following specification:
100V-240V~, 50Hz/60Hz.
WARNING. If the protective grounding (protective earth) system is in
doubt, the power of the system must be supplied by internal power
supply only.
To disconnect from the mains, the plug must be removed. Always
ensure that the power switch and plug are easily accessible.
NOTE. If the mains supply is interrupted and a battery is installed, the device
will switch to internal power supply and operate normally. If a battery is not
installed, the system will shut down and resume the previous settings at the
subsequent operation.
NOTE. After the AC power supply is connected, wait for at least 2 seconds
before pressing the POWER switch to turn on the system.
2.2.5 Configuration
The system can support, at most,
• 2 US-T transducers and 1 TOCO-T transducer
• 2 US-T transducers and 1 TOCO-E transducer (DECG and MHR not enabled),
• 1 US-T transducer and 1 TOCO-E transducer connected with DECG cable,
• 2 US-T transducers and 1 TOCO-E transducer connected with MECG cable.
Please do not exceed the maximum number.