41
Instructions For Use
EN
Enclosure Leakage
Current (Limit)
N.C. S.F.C.
100µA 500µA
Patient Leakage
Current (Limit)
FHR1, FHR2, TOCO
N.C. S.F.C.
d.c. 10µA 50µA
a.c. 100µA 500µA
Patient Auxiliary
Current (Limit)
FHR1, FHR2, TOCO
N.C. S.F.C.
d.c. 10µA 50µA
a.c. 100µA 500µA
Patient Leakage
Current (Limit)
DECG, MHR(from MECG)
N.C. S.F.C.
d.c. 10µA 50µA
a.c. 10µA 50µA
Patient Auxiliary
Current (Limit)
DECG, MHR(from MECG
N.C. S.F.C.
d.c. 10µA 50µA
a.c. 10µA 50µA
Base Station
Weight
About1.8 kg
Size
310mm x 235mm x81mm
US-T Transducer
Weight
About 150 g
Size
Ø81 mm × 35 mm
TOCO-T Transducer & TOCO-E Transducer
Weight
About 150 g
Size
Ø81 mm × 35 mm
A1.3 Performance Specifications
Ultrasound
*FHR Measurement Range
50 bpm ~ 240 bpm
*Resolution
1 bpm
*Accuracy
±2 bpm
Technique
Ultrasound Pulse Doppler with autocorrelation
Pulse Repetition Rate
2 kHz
Pulse Duration
92 µs
Ultrasound Frequency
(1+10%) MHz
p_
< 1 MPa
I
ob
<10 mW/cm2
I
spta
<100 mW/cm2
Dielectric Strength
4000 Vrms