34
Instructions For Use
EN
3. Maintenance and Cleaning
3.1 Maintenance
WARNING. All corrective maintenance must be performed by qualified
engineers, approved by Huntleigh.
The Sonicaid FTS-3 Service manual (PN 777882) is designed as an aid to engineers in
maintenance and service of repairable parts.
3.1.1 Maintaining Inspection
1. Visual Inspection
Prior to using FTS-3, do the following inspections:
• Check the system and accessories to see if there is any visible evidence of
damage that may affect patient safety. Pay special attention to any cracks on the
transducers and cables before immersing them into conductive fluid.
• Check all the outer cables, power socket and power cables.
• Check if the system functions properly
If any damage is detected, stop using the system on the patient. Replace the damage
part(s) or contact the manufacturer for service before reusing it.
2. Routine Inspection
The overall check of the base station and the accessories, including safety check and
function check, should be performed by qualified personnel every 6 to 12 months, and
each time after service.
The equipment should undergo periodic safety testing to ensure proper patient
isolation from leakage currents. This should include leakage current measurement and
insulation testing. The recommended testing interval is once a year or as specified in
the institution’s test and inspection protocol.
3. Mechanical Inspection
Make sure all exposed screws are tight.
Check the external cables for splits, cracks or signs of twisting.
Replace any cable that shows serious damage.
Pay particular attention to the supply socket.
WARNING. Failure on the part of the responsible individual hospital
or institution employing the use of this equipment to implement a
satisfactory maintenance schedule may cause undue equipment failure
and possible health hazards.
CAUTION. Besides the maintenance requirements recommended in this
manual, comply with local regulations on maintenance and measurement.