43
Instructions For Use
EN
MHR
* MHR Alarm Limits
30 bpm ~ 240 bpm
*Anti-electric Shock Type
Defibrillating-proof
ECG falls off
Detect automatically
Patient Leakage Current
(Limit)
N.C. S.F.C.
d.c. 10µA 50µA
a.c. 10µA 50µA
Patient Auxiliary Current
(Limit)
N.C. S.F.C.
d.c. 10µA 50µA
a.c. 10µA 50µA
Differential Input Impedance
>5MΩ
Electrode Offset Potential
Tolerance
±500mV
Auxiliary Current (Leads off
detection
Active electrode: < 100 nA
Reference electrode < 900 nA
Accuracy and Response to
Irregular Rhythm
Not Supported
Response time to Change in
MHR
MHR range:
80bpm ~
120bpm
Range: 7s~11s (average:9s)
MHR range:80bpm ~ 40bpm
Range : 8s ~ 12s (average: 10s)
Tall T-wave Rejection
Exceeds ANSI/AAMI EC13-2002
Sect. 3.1.2.1 (C)
maximum recommended 1.5mV
T-Wave amplitude
AFM
*Display Range
0 ~ 999
*FM Mode
Automatic
*AFM Mode
Trace (default) or Black Mark
Technique
Pulsed Doppler ultrasound
NOTE: The essential performance is marked with an asterisk *.
A1.4 Rechargeable Lithium-ion Battery
Base Station Battery
Nominal Capacity
5000 mAh
Operating time
≥ 40 Hours
Nominal Voltage
14.8 V
Charging time
≤ 14 Hours
Cycle Life
>300 times