37
Instructions For Use
EN
CAUTION. Please do not light the TOCO transducer with ultraviolet light for a
long time.
NOTE. The manufacturer has no responsibility for the effectiveness of
controlling infectious disease using these chemical agents. Please contact
infectious disease experts in your hospital for details.
In order to avoid infection, clean and disinfect the base station and accessories after
each use.
The recommended cleansers for accessories are listed below:
Accessory
Cleansers
Wireless US and TOCO Transducers
Mild near neutral detergent
Ethanol 75%
Isopropanol 70%
DECG Leads
ECG Leads
CAUTION.
• Be sure the temperature of cleaning solutions does not 45 ºC
(+113ºF).
• Only wipe the outer surface of accessories. Do not immerse them in any
liquid.
• Make sure no liquid enters the connector.
• After cleaning, no remaining cleanser is allowed on the surface.
• Clean the charging point periodically or it will not be charged.
3.2.1 Cleaning the Base Station
Regular cleaning of the base station enclosure is strongly recommended.
The solutions recommended for system cleaning are: mild near neutral detergent,
ethanol 75% and isopropanol 70%.
Clean the base station enclosure with a soft cloth and diluted non-caustic detergents
recommended above.
Clean the charging point in the docking slot with a dry soft cloth.
3.2.2 Cleaning the Transducers and Leads
Cleaning the Transducers
Make sure that there is no residual coupling gel.
The transducers must be thoroughly cleaned and disinfected at least once a month.
When cleaning, please firstly use a lint-free cloth moistened with mild near neutral
detergent, ethanol 75% solution or isopropanol 70% alcohol-based solution to clean
the transducers. Wipe using a cotton cloth moistened with water.
Use a dry, soft cloth to dry them.