32
Instructions For Use
EN
6.
Remove the TOCO-T or TOCO-E transducer and test the following
function:
•
The TOCO-T or TOCO-E transducer screen displays the standard start
interface.
•
The TOCO-T or TOCO-E transducer indicator is green.
•
The monitor screen displays TOCO.
7.
Touch the measuring area of the TOCO-T or TOCO-E transducer gently:
•
The fetal monitor displays TOCO value change.
8.
Install the US-T transducer to charge:
•
The US-T transducer screen displays charging interface and charging state.
•
The US-T transducer indicator is off.
•
The fetal monitor screen is blank.
9.
Install the TOCO-T or TOCO-E transducer to charge:
•
The TOCO-T or TOCO-E transducer screen displays charging interface and
charging state.
•
The TOCO-T or TOCO-E transducer indicator is off.
•
The fetal monitor screen is blank.
10.
It will take about 3.5 hours to charge the US-T, TOCO-T or TOCO-E
transducers.
2.3.8 Performance Test
To test a US transducer:
1 Connect the FTS-3 system to the monitor and switch it on.
2 Take up a US-T transducer and make sure it is successfully connected to the base
station.
3 Hold the transducer with one hand, and gently touch the center of the transducer
with the other hand in the frequency of 2 times per second.
4 Check that the value on the display shows this change in FHR.
To test a TOCO transducer:
1 Connect the FTS-3 system to the monitor and switch it on.
2 Take up the TOCO-T transducer or the TOCO-E transducer. Make sure it is
successfully connected to the base station.
3 Gently press the center of the transducer.
4 Check that the value on the display shows this change in pressure.
If a transducer fails the test, repeat this test with another transducer. If the second one
passes the test, defect of the first transducer is confirmed. Replace it with a good one.
If the second transducer fails the test as well, contact the manufacturer for service.