56
Instructions For Use
EN
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless
communications equipment
Test
Frequency
(MHz)
Band
a
(MHz)
Service
a
Modulation
b
Maximum
Power(W)
IMMUNITY
TEST LEVEL
(V/m)
385
380-390
TETRA 400
Pulse modulation
b
18Hz
1.8
27
450
30-470
GMRS 460, FRS
460
FM
C
±5 kHz deviation
1kHz sine
2
28
710
704-787
LTE Brand 13, 17
Pulse modulation
b
217 Hz
0.2
9
745
780
810
800-960
GSM
800/900,TETRA
800, iDEN 820,
CDMA 850, LTE
Band 5
Pulse modulation
b
18 Hz
2
28
870
930
1720
1700-1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE Band
1, 3, 4,25; UMTS
Pulse modulation
b
217 Hz
2
28
1845
1970
2450
2400-2570
Bluetooth,
WLAN,802.11
b/g/n, RFID 2450,
LTE Band 7
Pulse modulation
b
217 Hz
2
28
5240
5100-5800
WLAN 802.11
a/n
Pulse modulation
b
217 Hz
0.2
9
5500
5785
Note: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna
and the ME EQUIPMENT or ME SYSTEM maybe reduce to 1m. The 1 m test distance is permitted by IEC
61000-4-3.
a
For some services, only the uplink frequencies are included.
b
The carrier shall be modulated using a 50% duty cycle square wave signal.
c
As an alternative FM modulation, 50% pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.