Note: The specified accuracy is valid for all test results when Zero is performed for
each test (One Time Zero function is switched off).
2.4.6
User Options
User Method List
After switching on the instrument, a scroll list of all available methods is
automatically shown in the display. To shorten this list according to the
requirements of the user, it is possible to create a user defined scroll list.
The programme structure requires that this list must have at least one active
(switched on) method. For this reason, it is necessary to activate first all require
methods and then to switch off the automatically activated one if this method is not
required.
User Method List, Adaptation
Press [MODE], [Shift] + [6] [0] keys.
Confirm with [ ] key.
The display shows:
Start with [ ] key.
The complete method list is displayed:
Method with a point [•] behind the method
number will be displayed in the method selection
list. Methods without a point will not be displayed
in the method selection list.
Press arrow key [ ] or [ ] to select the required
method from the displayed list.
<Method list>
selected: •
Toggle: F2
save:
cancel: ESC
<Method list>
>>
20
•Acid demand
35•Alkalinity-p T
30•Alkalinity-tot T
….
Содержание WSL50 Pro
Страница 1: ...AquaSafe WSL50 Pro Water Safety Laboratory Instruction Manual ...
Страница 4: ...Section 2 Kit Contents c c c c b a e f d ...
Страница 7: ...Section 3 ...
Страница 8: ...AquaSafe Micro biological Incubator Instruction Manual ...
Страница 10: ...6 0 Guarantee and Assistance 32 ...
Страница 12: ......
Страница 16: ......
Страница 21: ......
Страница 33: ...Section 4 ...
Страница 34: ...AquaSafe Membrane Filtration Instruction Manual ...
Страница 50: ...Section 5 HydroTest Photometer HT1000 ...
Страница 86: ......
Страница 136: ......
Страница 142: ...HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Страница 193: ...NOTES ...