Place the vial in the sample chamber, making sure that the marks on the
instrument and vial are aligned.
Press the Test key
Remove the vial from the sample chamber.
Add one HT16 DPD No.3 tablet straight from the foil to the water
sample.
Crush the tablet using a clean stirring rod, until no large pieces are
visible.
Close the vial tightly with the cap and swirl several times until the tablet
is dissolved.
Place the vial in the sample chamber, making sure that the marks on the
instrument and vial are aligned.
Press the Test key
Wait for a reaction period of 2 minutes
After the reaction period, the measurement starts and the result is
displayed in:
mg/l free Chlorine
mg/l combined Chlorine
mg/l total Chlorine
LOD/Tolerance
The Lower LOD is 0.01 mg/l (10ppb), upper LOD is 6 mg/l (6000ppb).
Tolerance: ± 0.005mg/l.