Sample preparation (Free Chlorine):
Add one HT15 DPD No.1 tablet straight from the foil to the water
sample.
Crush the tablet using a clean stirring rod, until no large pieces are
visible.
Close the vial tightly with the cap and swirl several times until the tablet
is dissolved.
Analysis (Free Chlorine):
Place the vial in the sample chamber, making sure that the marks on the
instrument and vial are aligned.
Press the Test key
Wait for a few seconds until the result is displayed in mg/l free Chlorine.
Sample preparation (Total Chlorine):
Add one HT15 DPD No.1 tablet and one HT16 DPD No. 3 tablet straight
from the foil to the water sample.