4. Press ▲ or ▼ key to set the year.
5. Press ENTER key to confirm, the meter shows the currently programmed
date and time. The value currently being amended will be flashing.
6. Press ▲ or ▼ key to set each value
7. Press ENTER to cycle through to set each value.
8. When finished, press ENTER key to confirm; and the meter returns to
measurement mode.
4.5 Temperature Compensation
For accurate results, either manual or automatic temperature compensation
must be enabled before measurement or calibration
.
4.6 Automatic Temperature Compensation:
Insert the 6-pin connector on the temperature probe into the
corresponding connector on the instrument, ensuring that the lugs are
aligned. Once pushed firmly into place, tighten the waterproof locking
collar.
The “ATC” indicator will show on the display; the
meter is now
Enter
Enter